# TENDER FOR THE SUPPLY AND INSTALLATION of a Vector Network Analyser System covering frequencies up to 1.1 THz

[University of Birmingham](https://tendernation.io/buyers/university-of-birmingham)

Estimated value

£900,000 to £1,100,000

Published

14 Jul 2017

Deadline

Closed

14 Aug 2017

## Who can bid

- Buyer says suitable for SMEs

    The buyer ticked “suitable for SMEs” on the notice. Anyone can still bid.
- Open procedure

    Anyone can submit a full tender; there is no shortlisting stage.

[More contracts for small businesses and VCSEs](https://tendernation.io/opportunities/for-small-businesses)

## About this tender

The University of Birmingham invites tenders for the supply of the following equipment:

### THz Vector Network Analyser System

The system shall be capable of measuring, to a high degree of accuracy, the full two-port scattering parameters, (both magnitude and phase) of microwave devices in the frequency range 10 MHz to 1.1 THz.

The system shall include a base vector network analyser system covering 10 MHz to a suitable upper frequency (Band 1). The base VNA shall be usable on its own via a coaxial interface to cover Band 1, and usable in conjunction with a series of frequency converter units covering the waveguide bands from 140 GHz to 1.1 THz. Any gap in coverage between the top end of Band 1 and the lower frequency limit of the waveguide band converters should be kept to a minimum.

The frequency converter units, supplied in pairs to permit full two-port measurements, shall extend the frequency range over which full two-port measurements can be performed, to the following specific bands, and the measurement ports shall use the rectangular waveguide standards listed for each band:

o Band 2. 140 - 220 GHz, WR-5.1

o Band 3. 220 - 330 GHz, WR-3.4

o Band 4. 325 - 500GHz, WR-2.2

o Band 5. 500 - 750GHz, WR-1.5

o Band 6. 750 -1100 GHz, WR-1.0

The frequency converter heads shall meet the following requirements:

o Self-contained thermal management - no additional measures are needed to cool them.

o Individually characterised by the manufacturer in terms of conversion efficiency and output power.

A precision calibration kit shall be provided for each of the bands 1 to 6

The interface between the base VNA and the frequency converter units will be either built in to the base VNA or provided by a separate test set controller module, and shall provide the following:

o Power to drive the frequency converters.

o Sufficient length in the interface cables to allow at least 10 metres separation between two test ports, for antenna and propagation measurements.

o Capability to set up the base VNA to measure devices up to 1.1 THz, controlling up to four converter heads simultaneously.

o Switchable between VNA test ports and converter mode. The mm-wave modules and cables do not have to be removed to use the standard VNA from test ports for use in Band 1.

o Capability for the VNA to operate as a spectrum analyser in all bands up to 1.1 THz.

o Capability for different converter heads to be used on test port 1 and test port 2, or for one port to use the base VNA frequency while the other uses one of the frequency converters, to enable measurements to be performed on test devices with built in frequency conversion.

The base VNA shall meet the following requirements:

o A user friendly user interface

o Sufficient storage capacity for calibration data to avoid the need for frequently repeated calibrations by multiple users.

o Coverage from 10 MHz to a high upper frequency limit (Band 1), ...

Additional information: The University of Birmingham invites tenders for the supply of the following equipment:

### THz Vector Network Analyser System

The system shall be capable of measuring, to a high degree of accuracy, the full two-port scattering parameters, (both magnitude and phase) of microwave devices in the frequency range 10 MHz to 1.1 THz.

The system shall include a base vector network analyser system covering 10 MHz to a suitable upper frequency (Band 1). The base VNA shall be usable on its own via a coaxial interface to cover Band 1, and usable in conjunction with a series of frequency converter units covering the waveguide bands from 140 GHz to 1.1 THz. Any gap in coverage between the top end of Band 1 and the lower frequency limit of the waveguide band converters should be kept to a minimum.

The frequency converter units, supplied in pairs to permit full two-port measurements, shall extend the frequency range over which full two-port measurements can be performed, to the following specific bands, and the measurement ports shall use the rectangular waveguide standards listed for each band:

o Band 2. 140 - 220 GHz, WR-5.1

o Band 3. 220 - 330 GHz, WR-3.4

o Band 4. 325 - 500GHz, WR-2.2

o Band 5. 500 - 750GHz, WR-1.5

o Band 6. 750 -1100 GHz, WR-1.0

The frequency converter heads shall meet the following requirements:

o Self-contained thermal management - no additional measures are needed to cool them.

o Individually characterised by the manufacturer in terms of conversion efficiency and output power.

A precision calibration kit shall be provided for each of the bands 1 to 6

The interface between the base VNA and the frequency converter units will be either built in to the base VNA or provided by a separate test set controller module, and shall provide the following:

o Power to drive the frequency converters.

o Sufficient length in the interface cables to allow at least 10 metres separation between two test ports, for antenna and propagation measurements.

o Capability to set up the base VNA to measure devices up to 1.1 THz, controlling up to four converter heads simultaneously.

o Switchable between VNA test ports and converter mode. The mm-wave modules and cables do not have to be removed to use the standard VNA from test ports for use in Band 1.

o Capability for the VNA to operate as a spectrum analyser in all bands up to 1.1 THz.

o Capability for different converter heads to be used on test port 1 and test port 2, or for one port to use the base VNA frequency while the other uses one of the frequency converters, to enable measurements to be performed on test devices with built in frequency conversion.

The base VNA shall meet the following requirements:

o A user friendly user interface

o Sufficient storage capacity for calibration data to avoid the need for frequently repeated calibrations by multiple users.

o Coverage from 10 MHz to a high upper frequency limit (Band 1), to limit the gap in the overall frequency coverage of the system below 140 GHz.

o Sufficient output power to drive the converter heads, even when using cables for antenna and propagation measurements over a range of up to 10 m.

o System Dynamic range &gt;110dB.

Is a Recurrent Procurement Type? : No

Read full description

## CPV classification

Laboratory, optical and precision equipments (excl. glasses)

Source

Contracts Finder (UK)

[View official notice](https://www.contractsfinder.service.gov.uk/notice/4826a68e-a9d3-4a4c-b0ae-ea3fb9d8bc89)

Never miss a tender like this again

Get AI summaries, deadline alerts and a bid pipeline for tenders matched to your business. Free to start.

[Track tenders free](https://tendernation.io/register)

[Buyer profile University of Birmingham: past awards and live tenders](https://tendernation.io/buyers/university-of-birmingham) [Sector More Laboratory & precision equipment tenders](https://tendernation.io/sectors/laboratory-precision-equipment)

## More tenders like this

[Laboratory, optical and precision equipments (excl. glasses) – Dostawa aparatury naukowo – badawczej dla INTiBS PAN we Wrocławiu, z podziałem na dwa zadania: ZADANIE 1: stacjonarny analizator impedancji (0,01 Hz – 1 MHz) ZADANIE 2: stacjonarny analizator impedancji (1MHz - 1GHz) · Instytut Niskich Temperatur i Badań Strukturalnych im. Włodzimierza Trzebiatowskiego Polskiej Akademii Nauk · 15 days left · 3 Nov 2026](https://tendernation.io/opportunities/01a156a9-619a-714a-bd19-e119865a29ce/poland-laboratory-optical-and-precision-equipments-excl-glasses-dostawa) [Laboratory, optical and precision equipments (excl. glasses) – Modernizácia vedecko-výskumnej infraštruktúry Katedry jadrovej chémie Prírodovedeckej fakulty Univerzity Komenského v Bratislave · Univerzita Komenského v Bratislave · 30 days left · 18 Nov 2026](https://tendernation.io/opportunities/01a156a9-65a9-72ca-9922-4b240629b051/slovakia-laboratory-optical-and-precision-equipments-excl-glasses-modern) [Laboratory, optical and precision equipments (excl. glasses) – Liquid Handling System · Helmholtz Zentrum München Deutsches Forschungszentrum für Gesundheit und Umwelt (GmbH) · 31 days left · 19 Nov 2026](https://tendernation.io/opportunities/01a156a9-9882-716e-89c9-cdf421537c1b/germany-laboratory-optical-and-precision-equipments-excl-glasses-liquid) [Detection and analysis apparatus – Zaawansowana platforma do badań emisji akustycznej. Znak sprawy: DZPZ.282.65.2026.MK · Politechnika Śląska · 31 days left · 19 Nov 2026](https://tendernation.io/opportunities/01a156aa-10e7-7353-9fc0-d245bd78322b/poland-detection-and-analysis-apparatus-zaawansowana-platforma-do-badan-emis) [Electron microscopes – Prístroje a zariadenia pre projekt: Vývoj a výroba udržateľných bio-vláknových uhlíkových anód pre vysokovýkonné lítium-iónové batérie na podporu dekarbonizácie a energetickej nezávislosti Košíc a Košického kraja · Technická univerzita v Košiciach · 30 days left · 18 Nov 2026](https://tendernation.io/opportunities/01a156a9-651c-7266-b85f-cef62e6c451c/slovakia-electron-microscopes-pristroje-a-zariadenia-pre-projekt-vyvoj-a-vy) [Vibration analysers – Dostawa wibrometru laserowego na potrzeby Politechniki Lubelskiej. · Politechnika Lubelska · 4 days left · 23 Oct 2026](https://tendernation.io/opportunities/01a156a9-7c9f-71ac-ba41-52e17a7ec92f/poland-vibration-analysers-dostawa-wibrometru-laserowego-na-potrzeby-politec)